Hydro-Flow Flowmeters
Hydroflow
Model 1100 Specifications
|
Mechanical
Specifications
|
|
Type
|
Full
Bore, Inline
|
|
Measurable
Fluids
|
Water;
Water-based Mixtures
|
|
Pipe
Sizes
|
1,
1.5, 2, 2.5, 3 in. (25 to 80 mm)
|
|
Fluid
Temperature
|
32
to 140°F (0 to 60°C)
|
|
Fluid
Pressure
|
Varies
by fitting. See Graph.
|
|
Ambient
Temperature
|
–
20 to 140 °F (– 29 to 60 °C)
|
|
Flow
Range
|
0.5
feet, or 0.15 meters, per second minimum 15 feet, or 4.5
meters, per second maximum
|
|
Measuring
Units
|
English
|
Gallons
|
|
Metric
|
Cubic
Meters
|
|
Other
|
Other
measuring units available upon request or measuring
units can be reconfigured using Hydro-Flow's Field-Pro,
PC compatible configuration software.
|
|
Accuracy
(Combined Linearity & Repeatability)
|
±0.5%
of full scale
|
|
Wetted
Parts
|
Vortex
Sensor
|
Ultem®
(Plastic)
|
|
Shedder
Bar
|
316
Stainless Steel
|
|
Flowmeter
Body
|
PVC
Schedule 80
|
|
Stem
|
Aluminum
|
|
O-rings
|
EPDM
|
|
Mounting
Options
|
PVC
Schedule 80 Tee Fitting or Flange
|
|
Straight
Run Piping
|
Typical
10 diameters upstream, 5 diameters downstream
|
|
Electrical
Specifications
|
|
Enclosure
|
Reinforced
Polycarbonate, NEMA 6
|
|
European
CE Mark
|
Approved
|
|
Output
Signal Options
|
Pulse
Output
|
Frequency
proportional to flow rate or pulse proportional .
Maximum pulse width is 5 ms; For other pulse widths, use
the Fludiyne’s Relay Output Module. See Measurable
Flow Rates for standard output scaling.
|
|
Analog
Output
|
4-20
mA analog current loop, current proportional to flow
rate. 4 mA = zero flow; 20 mA = maximum flow listed in
Measurable Flow Rates.
|
|
No
Output
|
Display
only. Power Supply:
|
|
Power
Supply
|
Pulse
Output
|
10-32
VDC power supply with current limited by series
resistance to between 5 and 20 mA.
|
|
Analog
Output
|
10-32
VDC compliance.
|
|
No
Output
|
8-32
VDC, 4 mA maximum.
|
|
Display
Option
|
LCD
display alternately shows 4-digit rate and 8-digit total
flow.
|
|