Products       That Work ........ Smart
Home | Site Map | Sales Contact | Lighting | Agriculture | Turf | Golf | Pivot Alert | Pools |  Search | Register | Order Now | Credit Ap
 

 
   
 


 

 


Hydro-Flow Flowmeters

Hydroflow Model 2300 Specifications

Mechanical Specifications

Type

Insertion

Measurable Fluids

Ultrapure water; deionized water; acids; solvents; water

Pipe Sizes

0.5 to 8 in. (15 to 200 mm)

Fluid Temperature/Pressure

Varies by fitting.

Ambient Temperature

– 20 to 140 °F (– 29 to 60 °C)

Flow Range

0.5 feet, or 0.15 meters, per second minimum 20 feet, or 6 meters, per second maximum

Measuring Units

English

Gallons

Metric

Cubic Meters

Other

Other measuring units available upon request or measuring units can be reconfigured using Hydro-Flow's Field-Pro, PC compatible configuration software.

Accuracy
(Combined Linearity & Repeatability)

± 1.0% of full scale

Wetted Parts

Sensor/Bar/Stem

PVDF

O-rings

Viton®

Mounting Options

Retrofit

Fits existing +GF+ Signet tee fitting. See Signet Retrofit Compatibility

Union Tee Fitting

PVDF or PP for 0.5" to 1.5"

Tee Fitting

CPVC or PVC for 0.5" to 1.5"

Wafer

CPVC, PVC, PVDF or PP for 2" to 8"

Straight Run Piping

Typical 10 diameters upstream, 5 diameters downstream


Electrical Specifications

Enclosure

Reinforced Polycarbonate, NEMA 6

European CE Mark

Approved

Output Signal Options

Pulse Output

Frequency proportional to flow rate or pulse proportional . Maximum pulse width is 5 ms; For other pulse widths, use the Fludiyne’s Relay Output Module. See Measurable Flow Rates for standard output scaling.

Analog Output

4-20 mA analog current loop, current proportional to flow rate. 4 mA = zero flow; 20 mA = maximum flow listed in Measurable Flow Rates.

No Output

Display only. Power Supply:

Power Supply

Pulse Output

10-32 VDC power supply with current limited by series resistance to between 5 and 20 mA.

Analog Output

10-32 VDC compliance.

No Output

8-32 VDC, 4 mA maximum.

Display Option

LCD display alternately shows 4-digit rate and 8-digit total flow.

 

 

Send mail to webmaster lross@uicorp.net with questions or comments about this web site
Copyright © 2000 Underhill International Corporation
Last modified: December 06, 2007